| Technical Specifications |
|
|
|
| Dewpoint Sensor Element |
|
| Type: |
Hyper Thin Film high capacitance Al2O3 |
| Dewpoint range: |
XTR-100: -148°F to +68°F (-100°C to +20°C)
XTR-65 -85°F to +68°F (-65°C to +20°C) |
| Capacitance: |
15nF to 200nF |
| Accuracy: |
±5.5°F (±3°C) |
| Repeatability: |
±0.9°F(±0.5°C |
| Temperature Range: |
-14°F-158°F (-10°C to +70°C) |
Sample flow range:
(linear vel. @ 1atm): |
Static to 100 m/s |
| Storage temperature: |
-40°F to +176°F (-40°C to +80°C) |
| Calibration method: |
Autocal, sensor saturates at dewpoint above +68°F (+20°C). NIST/NPL traceable factory multipoint calibration available optionally |
| Temperature and Pressure Measurement |
| The instrument measures the sample temperature with a precision integrated circuit sensor. Units may be optionally fitted with a pressure sensor. |
| Electronics: |
|
| Input resolution: |
0.1°C(dp) |
| Indicators: |
3.5 digit LCD with custom legends |
| Engineering units: |
°C, °F, PPM, LBS H2O/mm scf, gm H2O/M3 |
| Controls |
3 push buttons, all settings stored in EPROM |
| Outputs: |
Analog: 4 - 20mA; Digital: RS-232 (bidirectional) |
| Alarms: |
The 4 - 20mA or the digital output may be used by an external device to operate relays |
| Isolation |
Sensor and case are isolated from the current loop and shunted with 33V transorbs. |
| Mechanical |
|
| Enclosure |
Stainless Steel
Weatherproof cover optionally available |
| Pressure operating range: |
Standard: 500 PSI (34 bar)
Optional: 5,000 PSI (340 bar) |
| Electrical connections: |
2.1 mm power jack with retainer |
| Mechanical connections: |
14 mm x 1.25 mm and 3/4 x 16 threads |
| Cable; |
Two conductor cable |
| Power requirements: |
10 to 33 VDC, the instrument draws 4-20mA depending on measured dewpoint. |
| Warranty: |
1 year |